Basics
Normal operation of the StedyCon microscope through a normal confocal scan with the Pollen slide. Additional books will cover other types of scans, e.g., Z-Stacks, Tiled scans, STED.
These documents are to reinforce, not replace, hands on training on the microscope. All users MUST recieve training from Cell Imaging Staff prior to use of the instrument.
Stop: This is a first draft in progress. The information is not/may not be accurate.
Power On
One Photo of the bottom of the laser box and the bottom rack - all lit up with numbers 1-6
One photo of the microscope power switch with 7
- Laser Box:
- the Laser Key to ON. The
laser box is normally in standby (There aretop two lightson:PowerPower(Red) and Standby/Readyand(Orange)twoarelightson,off: Emission and Scanning). (Color of lights….)
Istwo therelights, aEmission powerand switch or is that just the power strip?
- If the computer is off, press the button on the front of the box.
- If the monitor is off, there is a small power button on right bottom.
- Double click the StedyCon icon on the left side of the screen. This should bring up the Welcome Screen
The Microscope
Your first task will be to get your sample in focus and in the correct position. This is done with the microscope and through the controls on the microscope itself.
- Apply a drop of oil to the lens if needed (60X, 100X).
(The lenses areparafocal,parfocal, so you can start at lower magnificationsfirst.)first and work up to the 60X, 100X. - Place your slide with the coverslip down. Tilt the transmitted light arm back using the white housing at the top.
- Use the X-Y
controlledcontroller to get close toposition.position (below). - Set up the appropriate light path (below).
- Use the microscope to adjust focus and position (below).
BrightfieldXY Stage Control
Joystick with labels for fast and slow
The XY stage controller is the joystick on the air table. Press the top right button to set the mode to fast / coarse motion. Press the bottom right button to set the mode to slow / fine motion.
Bright-field Light Path and Controls
3/4 view of microscope with numbers for all the steps and arrows for the light path
Light path set to Eye and Epi shutter set to ---
Epi Control switch
Bright-field controls
- Verify that the lamp is on by putting your hand in the air path between the lamp and the condenser. If it is not, it may be switched off on the left panel or the intensity set to off.
If you cannot see anything in the eyepieces, check your light path again.
Fluorescent Light Path and Controls
3/4
Fluorescent filters
Beam block
Epi shutter button
Eyes and channel
- Iris: Should be pulled out.
- A.S. Two knobs
- ND4: Neutral Density Filter: should be pulled out for maximum intensity
- ND8: Neutral Density Filter: should be pulled out for maximum intensity
- Filter: Should be pushed in. If out, it will block the path. Can be used as a beam block.
- DAPI
- FITC
- TxRed
- D-F-T
XYZ ControlsFocus
The1 stageFocus XYknob control- isUp the/ joystick.Down, 2 Coarse, Medium, Fine.
Data Acquisition
StedyCon Welcome screen
Acquisition
On the welcomeStedyCon Welcome screen, press Start Up. The process will take a few minutes. Wait for the status to change from Starting Up... to Ready for Imaging.
Click on Startup. The start process will take a few minutes. The laser box Standby/Ready light will turn green. Eventually, the status will be Ready for Imaging.
Status and Interlocks
Interlocks tab with 5 green and 2 red
Click on the Status Info button and a Status window will pop up in a new tab.
Double click on the Interlock line in the STEDYCON State section. This will pop up a new tab with the Interlocks status.
There are 7 interlocks. The first 5 shouldmust be Green before starting a scan.
- Key: Is the key on the laser box on?
- Head:
- Interlock 1:
- TL arm & Condenser: There are two interlocks:
- TL arm: is the transmitted light arm (tower) forward?
- Condenser: is the dial set so that the brightfield path is blocked
- Beam path: On the front of the microscope, switch the path to L100
- Software: will still be red.
- Combined: will still be red.
The fluorescence source is not part of the interlocks. You still need to block its path to get a good image.
- Use the lever under the filter wheel to block the fluorscencefluorescence source.
- Set the fluorescence filter to ----.
Start New Session
Click on Start New Session. This will bring up the main window and the Define New Session dialog.
New Session Dialog
- Session Name:
- Sample Name:
- Objective Magnification:
- Objective NA:
- Choose Your Dyes:
- Dye Preset: pick from drop down
- Display Name
- Colormap
- Colormap in Overlay
- Load Settings: pull up settings from ?
- Let's Go
Scan Controls Overview
YouMain canwindow getwith numbers for areas
1- Scan Control
Scan Control section
Acquire
scan.
Auto:finish this session, with the option to begin a new session or quit.
Acquisition Mode:so that the photon counts are high enough (>70?) but do not saturate.
STED: Excitation2- Image Viewer
Image Viewer
the Z-position will pop up.
Channel3- Scan Parameters
Scan Parameters - with important ones opened up
Z-Stack:
Acquire z-slice and select z-stack: The z-slice can be useful in selecting the z-stack and in setting the laser power based on all the depths.
Lenses
| Magnification | Immersion | Numerical Aperature | Working Distance |
| 4X | Air | ||
| 10X | Air | ||
| 20X | Air | ||
| 40X - Not There | |||
| 60X | Oil | ||
| 100X | Oil |
Save Your Data
Gallery View
In the gallery, you can edit what data you want to save.
Select images by clicking on the check box beside them - or use the Select all button.
Selected images can be downloaded, or you can download all.
Alternately, when you click Finish Session, you will have the option to download or delete the session images.
Copy
Whereyour dodata theto imagesyour go?own
Your data ends up in two places:
- The C drive downloads folder as a .zip file named: yyyy-mm-dd session name.zip
- The S drive as a folder: yyyy-mm-dd session name
To copy your data, copy the S drive folder to your own hard drive.
What happens to the C drive folder?
What is our data clean up policy for this microscope?
Clean up
Take your sample off the microscope.
If you used oil on the objective, please use lens paper to remove the oil. The process should be taught during the hands on training.
Power Off
TheOne power off procedure is basically the reversephoto of the microscope power on.switch with number 1
One Photo of the bottom of the laser box and the bottom rack - all off up with numbers 2-6
FinishMicroscope:Session.
Normally, the computer is left on or put to sleep.