Zeiss Axioscan 7

Location: HCI Room 1470

CI-Axio7.jpg

The Zeiss AxioScan 7 is a fast and flexible slide scanner that digitizes specimens to produce high-quality virtual slides. In addition to brightfield slide scanning, the Axioscan 7 can scan up to five fluorescent channels using fast switching LED lights. Up to 100 slides can be loaded onto the scanner, making it suitable for large projects, including digital archives. Scanned images are saved as CZI files and can be viewed using Zeiss freeware Zen-lite or third-party image analysis programs.

Main Features of the Axioscan 7:

  • Holds up to 100 slides.
  • Able to image both bright-field (e.g. H&E slides) and up to nine-colour fluorescent samples. 
  • Image areas can be detected automatically or marked out, to form montages of the whole sample, or multiple samples on each slide
  • Setup wizard to guide users through optimising image acquisition and autofocus parameters.
  • Once setup, the system can be left for unattended automated scanning e.g. overnight

Suggested Applications:

  • Fixed slides.

Configurations:

The Axioscan 7 has a replaceable filter turret and can be switched between two light sources.  The 10 position filter cube turret is typically used with the Colibri 7 switchable LED source.  The fast filter wheel is typically used with the X-Cite Xylis LED source.

Configuration Light Source Filter Set
1 Colibri 7 Filter Cubes
2 X-Cite Xylis Fast Filter Wheel

Only Cell Imaging staff are permitted to switch between configurations.   Please schedule our time and plan accordingly.

 Light Sources:

Colibri 7:

The Colibri 7 has 6 LEDs with excitation filters to provide 7 excitation bands.  (One LED has a motorized filter to split it into two bands.)

Excitation Bands (There is one too many here)

Band Wavelength / Bandwidth (nm) Some typical fluorophores
UV 385/30

API, Hoechst 33342, Hoechst 33258, Alexa Fluor 350, Alexa Fluor 405, Indo-1, eBFP / BFP, eGFP (wt), True Blue

V 423/44

Pacific Blue, Lucifer Yellow, Alexa Fluor 433, eCFP, Cerulean

B 469/38

FM1-43, Cy2, eGFP, NBD, MitoTracker Green, Alexa Fluor 488, BCECF, Calcein, DiO SNAFL, YO-Pro-1, Nissl, LysoSensor Green, mHoneydew, FITC / Fluorescein, Kaede (green / red), PerCP, YoYo-1,FuraRed

C 511/44

Rhodamine 123, Fluo-4, Oregon Green BAPTA, Sytox Green, eYFP, FM4-64, Eosin/HE, Acridine Orange, JC1, Bodipy FL, Propidium Iodide, Spectrum Green, Calcium Green

G 555/30

TRITC, 7-AAD, Cy3, tdTomato, Alexa Fluor 546, Alexa Fluor 555, DsRed, mOrange, TagRFP, SNARF, DyLight 549, Spectrum Orange

Y 590/27

MitoTracker RED FM/CMXRos, txRed, mCherry, mRFP1, Cy3.5, Rhodamine B, Alexa Fluor 568, Dylight 594, Alexa Fluor 594, Bodipy TR

R 631/33

Alexa Fluor 633, Alexa Fluor 647, Cy5, DRAQ5, ToTo-3, ATTO-655, MitoTracker DeepRed, APC, ATTO-647N

FR 735/40

Alexa Fluor 750, Alexa Fluor 790, Cy7, Cy7.5

X-Cite Xylis:

The X-Cite Xylis XT720L is a broad spectrum illuminator with a range from 380-770nm.  It has peaks at 385, 430, 475, 545, 635, 735nm.

Filters:

Filter Cubes:

In multi-band fluorescence imaging, the filter turret has to move between positions for each band.  This is usually done on a per-tile basis.  At about 400ms per move, this can add up, especially for a large scan at higher magnification.  The Axioscan 7 has 2 multi-band cubes.  If these can be used, it can reduce or eliminate these moves and reduce scan time.  These can be used on the Axioscan 7 because it has the Colibri 7 switching LED source.

Filter No

 Fluorochrome

Excitation (nm)

Beamsplitter (nm)

Emission (nm)

1

Empty (Brightfield)

 

 

 

2

HE GFP

450-490

495

500-550

3

HE DsRed

538-562

570

570-640

4

Cy5

625-655

660

665-715

5

HE BFP

370-410

420

430-470

6

HE DAPI/GFP/Cy3.5/Cy7

375-395

455-483

583-600

720-750

405

493

611

762

410-440

501-547

618-650

770-800

7

HE DAPI/GFP/Cy3.5/Cy5/Cy7

370-400

450-488

540-570

614-647

720-750

405

493

575

654

761

412-438

501-527

582-601

662-700

770-800

Pol

Circular polarizer and analyzer




ND

6% Neutral Density Filter




Fast Filter Wheel:

The fast filter wheel moves between filter positions about 10 times faster than the filter cube turret.

Need information here

FilterPass Peak and WidthTypical fluorophores
Empty (Brightfield)

Single Channel 1

Single Channel 2

Single Channel 3

Dual Channel 4

Dual Channel 5

Objectives:

Objective Magnification Immersion

Numerical

Apperature

Correction Ring Coverglass (mm)

Working

Distance (mm)

Link
Fluar M27 5X Air 0.25
0.17

12.5


Pan Apo M27 10X Air 0.45
0.17 2.1
Pan Apo M27 20X Air 0.8
0.17 0.55
Pan Apo Korr M27 40X Air 0.95
0.13-0.21 0.25
EC Pan-Neofluar Pol M27 20X Air 0.5
0.17 2