# Zeiss Axioscan 7 #### Location: HCI Room 1470 [![CI-Axio7.jpg](https://bookstack.cores.utah.edu/uploads/images/gallery/2024-03/scaled-1680-/ci-axio7.jpg)](https://bookstack.cores.utah.edu/uploads/images/gallery/2024-03/ci-axio7.jpg) The Zeiss AxioScan 7 is a fast and flexible slide scanner that digitizes specimens to produce high-quality virtual slides. In addition to brightfield slide scanning, the Axioscan 7 can scan up to five fluorescent channels using fast switching LED lights. Up to 100 slides can be loaded onto the scanner, making it suitable for large projects, including digital archives. Scanned images are saved as CZI files and can be viewed using Zeiss freeware Zen-lite or third-party image analysis programs. #### Main Features of the Axioscan 7: - Holds up to 100 slides. - Able to image both bright-field (e.g. H&E slides) and up to nine-colour fluorescent samples. - Image areas can be detected automatically or marked out, to form montages of the whole sample, or multiple samples on each slide - Setup wizard to guide users through optimising image acquisition and autofocus parameters. - Once setup, the system can be left for unattended automated scanning e.g. overnight #### Suggested Applications: - Fixed slides. #### Configurations: The Axioscan 7 has a replaceable filter turret and can be switched between two light sources. The 10 position filter cube turret is typically used with the Colibri 7 switchable LED source. The fast filter wheel is typically used with the X-Cite Xylis LED source.
**Configuration****Light Source****Filter Set**
1Colibri 7Filter Cubes
2X-Cite XylisFast Filter Wheel
Only Cell Imaging staff are permitted to switch between configurations. Please schedule our time and plan accordingly. Light Sources: ##### Colibri 7: The Colibri 7 has 6 LEDs with excitation filters to provide 7 excitation bands. (One LED has a motorized filter to split it into two bands.) Excitation Bands (There is one too many here)
BandWavelength / Bandwidth (nm)Some typical fluorophores
UV385/30API, Hoechst 33342, Hoechst 33258, Alexa Fluor 350, Alexa Fluor 405, Indo-1, eBFP / BFP, eGFP (wt), True Blue
V423/44Pacific Blue, Lucifer Yellow, Alexa Fluor 433, eCFP, Cerulean
B469/38FM1-43, Cy2, eGFP, NBD, MitoTracker Green, Alexa Fluor 488, BCECF, Calcein, DiO SNAFL, YO-Pro-1, Nissl, LysoSensor Green, mHoneydew, FITC / Fluorescein, Kaede (green / red), PerCP, YoYo-1,FuraRed
C511/44Rhodamine 123, Fluo-4, Oregon Green BAPTA, Sytox Green, eYFP, FM4-64, Eosin/HE, Acridine Orange, JC1, Bodipy FL, Propidium Iodide, Spectrum Green, Calcium Green
G555/30TRITC, 7-AAD, Cy3, tdTomato, Alexa Fluor 546, Alexa Fluor 555, DsRed, mOrange, TagRFP, SNARF, DyLight 549, Spectrum Orange
Y590/27MitoTracker RED FM/CMXRos, txRed, mCherry, mRFP1, Cy3.5, Rhodamine B, Alexa Fluor 568, Dylight 594, Alexa Fluor 594, Bodipy TR
R631/33Alexa Fluor 633, Alexa Fluor 647, Cy5, DRAQ5, ToTo-3, ATTO-655, MitoTracker DeepRed, APC, ATTO-647N
FR735/40Alexa Fluor 750, Alexa Fluor 790, Cy7, Cy7.5
##### X-Cite Xylis: The X-Cite Xylis XT720L is a broad spectrum illuminator with a range from 380-770nm. It has peaks at 385, 430, 475, 545, 635, 735nm. #### Filters: ##### Filter Cubes: In multi-band fluorescence imaging, the filter turret has to move between positions for each band. This is usually done on a per-tile basis. At about 400ms per move, this can add up, especially for a large scan at higher magnification. The Axioscan 7 has 2 multi-band cubes. If these can be used, it can reduce or eliminate these moves and reduce scan time. These can be used on the Axioscan 7 because it has the Colibri 7 switching LED source.
**Filter No** **Fluorochrome** **Excitation (nm)** **Beamsplitter (nm)** **Emission (nm)**
1 Empty (Brightfield)
2 HE GFP 450-490 495 500-550
3 HE DsRed 538-562 570 570-640
4 Cy5 625-655 660 665-715
5 HE BFP 370-410 420 430-470
6 HE DAPI/GFP/Cy3.5/Cy7 375-395 455-483 583-600 720-750 405 493 611 762 410-440 501-547 618-650 770-800
7 HE DAPI/GFP/Cy3.5/Cy5/Cy7 370-400 450-488 540-570 614-647 720-750 405 493 575 654 761 412-438 501-527 582-601 662-700 770-800
Pol Circular polarizer and analyzer
ND 6% Neutral Density Filter
##### Fast Filter Wheel: The fast filter wheel moves between filter positions about 10 times faster than the filter cube turret. Need information here
FilterPass Peak and WidthTypical fluorophores
Empty (Brightfield)
Single Channel 1
Single Channel 2
Single Channel 3
Dual Channel 4
Dual Channel 5
#### Objectives:
**Objective****Magnification****Immersion****Numerical** **Apperature** **Correction Ring****Coverglass (mm)****Working** **Distance (mm)**
Fluar M275XAir0.25 0.1712.5
Pan Apo M2710XAir0.45 0.172.1
Pan Apo M2720XAir0.8 0.170.55
Pan Apo Korr M2740XAir0.95 0.13-0.210.25
EC Pan-Neofluar Pol M2720XAir0.5 0.172