# Leica SP8 405-488-561-633 Laser Confocal Location: HCI Room 1440 **Main features of Leica SP8 White Light System:** - Five detectors, Two Photo Multipliers Tube (PMT) detectors and three HyD detectors for enhanced sensitivity and dynamic range for 5 total fluorescent channels. - Transmitted light detector available for simultaneous brightfield imaging of samples while scanning confocally. - Hybrid confocal scan head unit can provide high resolution images up to 4096 x 4096 pixels in size in conventional galvonometer mode and in high speed resonant mode up to 30 frames per second at 512 x 512 pixels. - Ultra high speed resonant acquisitions up to 420 frames per second at 512 x 32 pixels. - Fully automated Leica DM8 microscope - Epifluorescence and DIC (Nomarski) illumination lamps for rapid sample identification and focusing. - Filter wheel used for epifluorescence viewing of sample is currently outfitted with with DAPI (blue), FITC (green), and TRITC (red), filter cube sets. - Laser autofocus system provides focal depth consistency across varying sample regions of interest. - OkoLab heating/cooling stage top incubator with C02 and humidity. Plates and dishes with glass bottoms. - z-galvo **Imaging lasers:** - 405nm diode laser - 488nm Argon gas laser - White Light continuously tunable pulsed laser 440-700nm - 561 - 633 **Suggested Applications:** - Multichannel fluorescence and transmitted fixed slide imaging - Tiling mosaics of fixed or live samples - Timelapse of dishes or glass bottom well plates #### Objectives:
**Objective** | **Magnification** | **Immersion** | **Numerical Aperture** | **Correction Collar** | **Coverslip (mm)** | **Working Distance (mm)** |
HC PL APO CS2 | 10 | Air | 0.4 | 2.74 | ||
HC PL APO CS2 | 20 | Either water, glycerine, or oil | 0.75 | Corr | 0.66 | |
HC PL APO CS2 | 20 | Air | 0.75 | 0.17 | 0.62 | |
HC PL APO CS2 | 40 | Water | 1.1 | Corr | 0.14-0.18 | 0.65 |
HC PL APO CS2 | 40 | Oil | 1.3 | 0.17 | 0.24 | |
HC PL APO CS2 | 63 | Oil | 1.4 | 0.17 | 0.14 |